Version 02.00, November 2014
Rohde & Schwarz R&S
®
BTC Broadcast Test Center 11
Phase coherence
The R&S
®
BTC provides phase-coherent RF outputs for the two RF paths or two or more instruments. For frequencies above 200 MHz,
the LO output and input frequency are set to the output frequency.
LO coupling modes
This mode corresponds to internal
LO operation in path A and path B.
A, B internal
This mode corresponds to internal
LO operation in path A, and LO of path B
is coupled to path A.
A internal,
A → B coupled
This mode corresponds to external
LO operation at the LO IN connector in
path A and internal LO operation in path B.
A external,
B internal
This mode corresponds to external
LO operation at the REF/LO IN connector
in path A and path B.
A external,
A → B coupled
REF/LO OUT states
The active LO signal of path B can be
routed to the LO OUT connector (in order
to couple two or more instruments).
on/off
Input of phase coherence signal
Connector type
LO IN on rear panel
SMA female
Input impedance
50 Ω (nom.)
Input level range of external LO signal
7 dBm to 13 dBm
Frequency range of external LO signal
for RF setting 200 MHz < f ≤ 6 GHz
200 MHz to 6 GHz
Output of phase coherence signal
Connector type
LO OUT on rear panel
SMA female
Output impedance
50 Ω (nom.)
Output level range of internal LO signal
7 dBm to 13 dBm
Frequency range of internal LO signal
for RF setting 200 MHz < f ≤ 6 GHz
200 MHz to 6 GHz